Sysmex XP-300 14-1
Technical Information
Revised March 2012
14. Technical Information
14.1 Specifications
Operating
Environment
Ambient temperature: 15°C to 30°C (same for the supplied reagent temperature)
Relative humidity: 30% to 85%
Atmospheric pressure: 70 - 106 kPa
Operating altitude: Maximum 3,000 m
Contamination Level 2
Impulse Resistance
(Excess voltage)
category
II
Usage location Indoor usage only
Storage
(transportation)
condition
Ambient temperature: -10°C to +60°C
Relative humidity: 30% to 95% (non condensation/keep dry)
Atmospheric pressure: 70 - 106 kPa
Analysis parameters WBC (Number of leukocytes), RBC (Number of erythrocytes),
HGB (Hemoglobin), HCT (Hematocrit value), MCV (Mean RBC volume),
MCH (Mean RBC hemoglobin), MCHC (Mean RBC hemoglobin concentration),
PLT (Platelet), RDW-SD (RBC distribution width - SD),
RDW-CV (RBC distribution width - CV), PDW (Platelet distribution width),
MPV (Mean platelet volume), P-LCR (Large platelet ratio), PCT (Plateletcrit),
LYM% (W-SCR), MXD% (W-MCR), NEUT% (W-LCR), LYM# (W-SCC),
MXD# (W-MCC), NEUT# (W-LCC)
Display range WBC 0.0 - 299.9 × 10
3
/µL
RBC 0.00 - 19.99 × 10
6
/µL
HGB 0 - 25.0 g/dL
PLT 0 - 1999 × 10
3
/µL
Analysis range WBC 1.0 - 99.9 × 10
3
/µL
RBC 0.3 - 7.0 × 10
6
/µL
HGB 0.1 - 25.0 g/dL
HCT 10.0 - 60.0 %
PLT 10.0 - 999 × 10
3
/µL
Background limits WBC 0.3 × 10
3
/µL or less
RBC 0.02 × 10
6
/µL or less
HGB 0.1 g/dL or less
PLT 10 × 10
3
/µL or less
Reagent Diluent: CELLPACK
WBC/HGB lyse reagent: STROMATOLYSER-WH
Detergent CELLCLEAN