4-2 Sysmex XP-300
Reagents
Revised February 2013
4.3 STROMATOLYSER-WH
STROMATOLYSER-WH is a reagent that lyses RBC for accurate
WBC count determination, WBC tri-modal size distribution analysis
and hemoglobin level measurement.
Caution!
Leave STROMATOLYSER-WH at room temperature
(15 - 30°C) for at least 24 hours or longer before
using. If a reagent that has arrived recently is used, a
correct analysis result may not be obtained.
Use STROMATOLYSER-WH at 15 - 30°C.
When analyzed with reagent of a temperature higher
than 30°C or lower than 15°C, a correct analysis
result may not be obtained.
Do not allow STROMATOLYSER-WH to freeze.
STROMATOLYSER-WH is solely intended for use
with Sysmex reagents and analyzers.
If other reagents are used, the product performance
of Sysmex instruments cannot be guaranteed.
Do not use any left-over STROMATOLYSER-WH.
Always use new STROMATOLYSER-WH.
Take care to prevent dust, dirt, bacteria, or other
materials from entering the reagent after it is opened.
Otherwise, there is a possibility that proper analysis
cannot be performed.
Do not use STROMATOLYSER-WH that shows
signs of contamination or instability, such as turbidity
or discoloration.
Information
Storage and shelf life after first opening
Store STROMATOLYSER-WH at 2 - 35°C.
The expiration date is shown on the outer packaging.
Once opened (connected to the instrument), product
stability of the reagent is a maximum of 90 days. In
this case, however, make sure to store
STROMATOLYSER-WH at 2 - 35°C.