Sysmex XP-300 12-1
Cleaning and Maintenance
January 2012
12. Cleaning and Maintenance
To ensure proper functioning of the instrument, it is necessary to
periodically clean and service the instrument. Perform maintenance
according to the schedule below. And record the results in the
Maintenance checklist.
12.1 Maintenance schedule
Daily
Clean TD chambers and diluted sample lines (Shutdown)
(see 12.3)
Check trap chamber level and discard (see 12.4)
Weekly
Clean SRV tray (see 12.5)
Every month (or every 1,500 samples)
Clean TD (see 12.6)
Clean waste chamber (see 12.7)
Every 3 months (or every 4,500 samples)
Clean SRV (see 12.8)
As-needed maintenance
Check instrument status (see 12.2)
Perform Auto Rinse (see 12.9)
Clean rinse cup (see 12.10)
Dispose waste fluid (see 12.11)
Clean aperture of TD chamber (see 12.12)
Calibrate LCD screen (see 12.13)
Reset SRV cycle counter (see 12.14)
Adjust pressure and vacuum (see 12.15)
Risk of Infection
To avoid the risk of infections, wear protective
garments and gloves for all cleaning or maintenance
work. After completion of work, wash hands with
disinfectant. Otherwise, there is a risk of infection by
pathogens.