Sysmex XP-300 12-11
Cleaning and Maintenance
Revised July 2012
12.8 Clean SRV
When the main power switch is turned ON, and if either the counter
value exceeds 4,500, or if 3 months have passed since the last
maintenance, a message will appear prompting the operator to
perform periodic maintenance (SRV cleaning).
Press the [Cancel] button to continue start up without executing
SRV cleaning operation. Until the SRV is cleaned, the above
message is displayed at start up.
When this message is displayed, perform the SRV cleaning
according to the following procedures.
(1) Press the [OK] button.
The SRV counter is reset, and the main power switch can be
turned OFF.
(2) Turn off the power of the main unit and wait approximately 30
seconds.
(3) Open the front cover of the main unit.
Risk of Infection
When cleaning the SRV, always wear rubber gloves.
After completion of work, wash hands with
disinfectant.
If your hands are contaminated by blood, etc.,
infection of bacteria can occur.
Warning!
CELLCLEAN is a strong alkaline cleaning solution. Do
not mix it with any acidic substances. Also, it should
not come in contact with skin or clothing. If the skin or
clothes should come in touch with it, flush it away
using plenty of water.
Otherwise, it can damage the skin or clothes.
Caution!
Before turning the power off, always press the [OK]
button, making the instrument ready to be turned off.
Note:
When cleaning the SRV before the above message is
displayed, reset the SRV counter (see “12.14 Reset SRV
cycle counter”).