14.2 System limits
Possible sample interferences
WBC
If any of the following is present, the system may erroneously report a low white blood cell count.
Leukocyte aggregation
If any of the following is present, the system may erroneously report a high white blood cell count.
Possibility of PLT clumps
Lyse resistant erythrocytes
Erythrocyte aggregation (cold agglutinin)
Cryoprotein
Cryoglobulin
Fibrin
Giant platelets (platelets > 1,000,000/µL)
RBC
Where the following are present, the system may erroneously report a low red blood cell count.
Erythrocyte aggregation (cold agglutinin)
Microerythrocytes
Possibility of fragmented RBCs
If any of the following is present, the system may erroneously report a high red blood cell count.
Leukocytosis (> 100,000/µL)
Giant platelets (platelets > 1,000,000/µL)
HGB
If any of the following is present, the system may erroneously report a high haemoglobin concentra-
tion.
Leukocytosis (> 100,000/µL)
Lipemia
Abnormal protein