EasyManua.ls Logo

SYSMEX XP Series - End of Operation (Shutdown)

SYSMEX XP Series
290 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-28 Sysmex XP-300
Sample Analysis
January 2012
7.12 End of operation (Shutdown)
At the end of the day’s analysis, before the instrument is turned OFF,
the Shutdown sequence should be executed. Executing a Shutdown
cleans TD chambers and diluted sample lines.
Executing a Shutdown:
To be carried out after all analysis have been performed, or at least
once every 24 hours, if instrument is used continuously.
(1) Press the [Shutdown] button at the Ready status.
The shutdown confirmation dialog screen will appear.
In case the [Cancel] button is pressed, the Shutdown is
canceled, and the instrument returns to the Main screen.
Caution!
If the instrument is used continuously without
performing the Shutdown sequence, protein build-up
on the internal parts may prevent obtaining correct
analysis results, and may damage the instrument.
Information
If the instrument is turned OFF without executing a
Shutdown, water droplets may come out of the rinse
cup or deposits may build-up on the rinse cup.
Note:
It takes approximately 5 minutes to complete the Shutdown
sequence.
If you repeatedly restart the instrument after Shutdown, the
shutdown confirmation dialog screen appears each time.

Table of Contents

Related product manuals