Sysmex XP-300 7-29
Sample Analysis
Revised July 2012
(2) Set the CELLCLEAN to the sample probe, and in that
condition, press the start switch.
While [Aspirating] is being displayed on the screen, keep
holding CELLCLEAN in the same status, as long as the “beep”
is present.
(3) Check that the shutdown sequence was completed and that the
shutdown completion screen is displayed.
Turn OFF the main power switch on the right side of the
instrument.
When the [Restart] button is pressed, the Main screen will
appear.
CELLCLEAN is a strong alkaline cleaning solution. Do
not mix it with any acidic substances. Also, it should
not come in contact with skin or clothing. If the skin or
clothes should come in touch with it, flush it away
using plenty of water. Otherwise, it can damage the
skin or clothes.
• When the sample is being aspirated, [Aspirating]
appears on the screen. Do not remove the
CELLCLEAN from the sample probe while
[Aspirating] is displayed. Otherwise, there is a
possibility that proper aspiration cannot be
performed.
• To remove CELLCLEAN, lower it straight down.
Take care not to bend the sample probe.