Sysmex XP-300 2-5
Safety Information
Revised February 2013
2.6 Maintenance
Always wear protective garment and gloves for all
service and maintenance work. Use only specified
tools and parts. After work is completed, wash hands
with disinfectant. Instrument surfaces in contact with
blood is potentially biohazardous.
Do not use any organic solvent, acid, or alkaline agent
(except CELLCLEAN that is 1:10 diluted with water).
These will affect the instrument surface's finish and
may cause corrosion or discoloration.
• Do not perform any maintenance work or repair other
than that as specified in this manual.
• When performing out maintenance work, use only
the specified tools and parts. Install only such spare
or replacement parts intended for this equipment.
• Use CELLCLEAN that is 1:10 diluted with water to
clean blood, etc. stained on the instrument surface.
When cleaning the touch panel surface, use a “soft
cloth dampened with ethanol”.
• If the instrument is not operated every day, a “Blank
Error” may occur.
In this case, follow the instructions displayed in the
action message screen.