2-2 Sysmex XP-300
Safety Information
Revised February 2013
2.2 Installation
2.3 Electromagnetic compatibility (EMC)
This instrument complies with the following IEC (EN) standards:
IEC61326-2-6:2005 (EN61326-2-6:2006)
Electrical equipment for measurement, control and laboratory
use - EMC requirements
EMI (Electromagnetic Interference) For this standard the
requirements of class A are fulfilled.
EMS (Electromagnetic Susceptibility) For this standard the
minimum requirements with regards to immunity are fulfilled.
This equipment has been designed and tested to CISPR 11 Class
A. In a domestic environment it may cause radio interference, in
which case, you may need to take measures to mitigate the
interference.
Do not use this device in close proximity to sources of strong
electromagnetic radiation (e.g. unshielded intentional RF
sources), as these may interfere with the proper operation.
Caution!
This instrument must be protected against splashing
water.
Install in places where there are no adverse effects
from high temperature and humidity, dust or direct
sunlight.
Avoid shock and vibrations.
The installation location must be well ventilated.
Avoid installation near devices causing potential
interference, such as wireless communication
equipment or similar devices, and centrifuges.
Installation of this instrument in places where
chemicals are stored or hazardous gas may be
present is not permitted.
Avoid installation near devices that are susceptible to
interference, such as personal computer monitors.
Use the instrument in places where ambient
temperature ranges between 15°C and 30°C
(optimal at 23°C).
Use the instrument in places where relative humidity
ranges between 30% and 85%.
Note:
Please always run the instrument with the front cover
closed. The instrument does not comply with the standards
mentioned above while the front cover is left open.