Sysmex XP-300 5-7
Initial Operation
Revised June 2013
5.5 Connection of reagents and waste bottle
Prepare reagent
Prepare the reagents and connection kits shown below.
Type Description Volume Connection Kit
PK-30L CELLPACK 20 L Cubitainer spout kit No.1 (for 20 L)
CPK-310A CELLPACK 10 L (a separate kit required)
SWH-200A STROMATOLYSER-WH 500 mL × 3 Float Switch No. 23 Assy
Caution!
Take care not to spill reagent on to the instrument. If it spills, wipe it off immediately using a
damp cloth.
If it comes into contact with your skin, wash it off using plenty of water.
If a reagent enters your eye, wash it out immediately using plenty of water, and seek medical
treatment at once.
After unpacking, be sure not to allow dust or dirt to come into the reagent. Use up CELLPACK
within 60 days from the unpacking date, and STROMATOLYSER-WH within 90 days.
Note:
When using the CELLPACK (10 L), a separate cubitainer spout kit is required. Contact your Sysmex
service representative.