6-6 Sysmex XP-300
Operation
Revised October 2012
6.3 Signal tones
This instrument indicates different situations by 4 distinct signal
tones.
(1) Key tone (Short beep)
Beeps for about 0.1 seconds when pressing a button which is
active input.
(2) Input errors (Long beep)
Beeps for about 1 second when pressing a button which is not
active input. Beep sounds (beep, beep, beep) when the
pneumatic unit is stopped.
(3) Analysis error
Continuously beeps if the instrument detects an error, and stops
the alarm sound when the [OK] button of the error dialog
displayed on the LCD screen is pressed. Settings can be used to
change the sound volume.
See “11. Instrument Setup”.
(4) Tone at sample aspiration
Usually: A single “beep” when the start switch is pressed;
“beep, beep” when aspiration is finished.
Sample No. “0” or in pre-diluted mode:
Beep sounds (beep, beep, …) continue from pressing
the start switch until aspiration ends.
6.4 Pneumatic unit stop function
Automatic stop function
This instrument, when left non-operating for 15 minutes, automatically
stops the pneumatic unit. This function prevents the hydraulic line
from becoming dry, saves power consumption and extends component
service life. In addition, with this function, analysis-ready status can be
resumed faster than by turning on the power.
Pressing the start switch, the instrument will return to “Ready” status.
Manual stop function
This instrument stops the pneumatic unit when the [PU Sleep] button
on the Menu screen is pressed.
Pressing the start switch, the instrument will return to “Ready” status.
Information
When an alarm is sounded, the alarm sound can be
stopped if the [OK] button of the error dialog displayed
on the LCD screen is pressed. Press the button
to display the help menu. All other buttons are not
functional during an alarm.