Sysmex XP-300 13-17
Troubleshooting
January 2012
9. Maintenance Errors
Error Message
(Error Code)
Clean the SRV (471010)
Cause 3 months have passed since the last maintenance or the number of samples
analyzed exceeds 4,500.
Action Troubleshooting
on the screen
Press the [OK] button.
Following the instructions on the screen, turn OFF the power
and clean the SRV. (Refer to “12.8 Clean SRV”).
If you perform cleaning later, press the [Cancel] button. This
activates the usual startup sequence and makes the system
Ready.
* The same message will appear the next time you start the
unit; it will be displayed at each startup until the cleaning is
performed.
Note If SRV cleaning is performed without pressing the [OK] button, reset the SRV
cycle counter by referring to “12.14 Reset SRV cycle counter”.
Error Message
(Error Code)
Clean Waste Chamber (471030)
Cause 1 month has passed since the last maintenance or the number of samples analyzed
exceeds 1,500.
Action Set CELLCLEAN to the sample probe and clean the waste chamber. (Refer to
“12.7 Clean waste chamber”.)
If you perform cleaning later, press the [Cancel] button. This activates the usual
startup sequence and makes the system Ready.
* The same message will appear the next time you start the unit; it will be
displayed at each startup until the cleaning is performed.
Note When the waste chamber rinse sequence has been completed, the cycle counter is
automatically reset.
Error Message
(Error Code)
Clean Transducer (471100)
Cause 1 month has passed since the last maintenance or the number of samples analyzed
exceeds 1,500.
Action Use CELLCLEAN to clean the WBC transducer and RBC transducer. (Refer to
“12.6 Clean TD”.)
If you perform cleaning later, press the [Cancel] button. This activates the usual
startup sequence and makes the system Ready.
* The same message will appear the next time you start the unit; it will be
displayed at each startup until the cleaning is performed.
Note When the TD rinse sequence has been completed, the cycle counter is
automatically reset.