13-16 Sysmex XP-300
Troubleshooting
Revised May 2012
8. Others
Error Message
(Error Code)
QC(L-J) Error (461160)
QC(X-bar) Error (461150)
Cause QC analysis results were beyond the control limit, or [+++.+], [∗∗∗.∗] or [---.-]
was displayed due to the following reason(s):
Control blood aspiration error
Control blood insufficient mixing
Faulty control blood
TARGET value or LIMIT value was input wrong.
Instrument failure
Action Reanalyze control blood.
Confirm TARGET value and LIMIT value.
Note Ready for analysis.
Error Message
(Error Code)
QC Error (466010)
Cause Input error of target or calibration values
Instrument failure caused data to shift.
Action Execute calibration again.
Press the [Back] button and execute calibration again from the beginning.
Maintenance of instrument
Check QC data, and if data is found to have shifted, a hardware problem is
conceivable.
Referring to “12. Cleaning and Maintenance”, clean the TD, SRV, etc.
Note Analysis is not ready. Not ready until calibration value is fixed.