13-16 Sysmex XP-300
Troubleshooting
Revised May 2012
8. Others
Error Message
(Error Code)
QC(L-J) Error (461160)
QC(X-bar) Error (461150)
Cause QC analysis results were beyond the control limit, or [+++.+], [∗∗∗.∗] or [---.-]
was displayed due to the following reason(s):
• Control blood aspiration error
• Control blood insufficient mixing
• Faulty control blood
• TARGET value or LIMIT value was input wrong.
• Instrument failure
Action • Reanalyze control blood.
• Confirm TARGET value and LIMIT value.
Note Ready for analysis.
Error Message
(Error Code)
QC Error (466010)
Cause • Input error of target or calibration values
• Instrument failure caused data to shift.
Action • Execute calibration again.
Press the [Back] button and execute calibration again from the beginning.
• Maintenance of instrument
Check QC data, and if data is found to have shifted, a hardware problem is
conceivable.
Referring to “12. Cleaning and Maintenance”, clean the TD, SRV, etc.
Note Analysis is not ready. Not ready until calibration value is fixed.