12-30 Sysmex XP-300
Cleaning and Maintenance
Revised February 2013
12.16 Replace reagent
During the operation, instrument will automatically stop and an error
dialog will appear if a reagent runs out.
To replace the reagent, follow the procedure below.
(1) Make ready a new reagent and make sure that it has not passed
its expiration date.
Always wear protective garments and gloves when
replacing reagents. After completion of work, wash
hands with disinfectant. Otherwise, there is a risk of
infection by pathogens.
Error message Reagent to be replaced
[Replace CELLPACK] CELLPACK
[Replace STROMATOLYSER] STROMATOLYSER-WH
• Leave the reagents at room temperature (15 -
30°C) for at least 24 hours or longer before using.
If a reagent that has arrived recently is used, a
correct analysis result may not be obtained.
• Use the reagent at 15 - 30°C. When analyzed with
reagent of a temperature higher than 30°C or lower
than 15°C, a correct analysis result may not be
obtained.
• Always use new reagents.
• Do not use any left-over reagents.
• Do not allow the reagents to freeze.
• Take care to prevent dust, dirt, bacteria, or other
materials from entering the reagent after it is
opened. Otherwise, there is a possibility that
proper analysis cannot be performed.
Shelf life after first opening
CELLPACK 60 days
STROMATOLYSER-WH 90 days