Sysmex XP-300 12-17
Cleaning and Maintenance
January 2012
12.10 Clean rinse cup
When the rinse cup has any blood on it or is clogged, clean it by the
following procedure:
(1) Turn off the power of the main unit and wait approximately 30
seconds.
(2) Open the front cover of the main unit.
(3) Gently push down the rinse cup using both hands.
Make sure the rinse cup is removed completely from the sample
probe.
(4) Remove the rinse cup in the order of (1), (2), and (3) as shown
on the left.
(5) Clean the rinse cup using tap water.
(6) Make sure no contaminant remains on the rinse cup and wipe
off water.
When cleaning the rinse cup, always wear rubber
gloves. After completion of work, wash hands with
disinfectant. If your hands are contaminated by blood,
etc., infection of bacteria can occur.