12-18 Sysmex XP-300
Cleaning and Maintenance
January 2012
(7) Mount the rinse cup in the reverse order of removal. Route the
thin tubes around the rear of the rinse cup.
(8) Gently push up the rinse cup to the top using both hands.
Make sure the rinse cup is pushed all the way up with
the sample probe inserted in the hole. If the power is
turned on with the rinse cup staying lowered, [Rinse
motor error] will occur, making it impossible to
continue the operation.