Sysmex XP-300 13-15
Troubleshooting
January 2012
7. Memory Errors
Error Message
(Error Code)
RAM Error (431010)
ROM Error (431020)
Cause Momentary power failure, sudden noise interference, etc. causes CPU
malfunction.
Action Turn OFF then ON the power again.
Note Applications cannot be started.
Notifies the operator of the error by an alarm sound because it cannot be displayed
on the screen.
Error Message
(Error Code)
Stored Error (432010)
Settings Error (431030)
QC Data Error (463000)
Cause Momentary power failure or sudden noise interference causes stored data/QC data/
set value errors.
Action Troubleshooting
on the screen
• Follow the instructions on the screen to “repair” the
data storage area. Then restart the instrument.
If “repair” is not successful, then “initialize” the data
storage area.
• When the instrument is initialized, all stored data is
deleted. If settings have been initialized, reenter the
settings.
• If either “repair” or “initialize” is successful, the
program will be started.
Note Analysis is not ready. Not ready until the error is resolved.