12-10 Sysmex XP-300
Cleaning and Maintenance
Revised July 2012
Cleaning waste chamber will be executed, and the cleaning
waste chamber screen will appear.
Then, Auto Rinse and background check will be performed.
(2) After completion of waste chamber cleaning, the Main screen
will appear.
• When analysis starts, the status reads [Aspirating].
When the status display [Aspirating] changes to
[Running], the CELLCLEAN container can be
removed safely.
Do not remove the CELLCLEAN container from the
sample probe while [Aspirating] is displayed.
Otherwise, there is a possibility that proper aspiration
cannot be performed.
• Several seconds after the buzzer sounds “beep,
beep” and [Running] appears on the screen, the
rinse cup lowers. Remove the CELLCLEAN
container by that time.
• To remove the CELLCLEAN container, lower it
straight down. Take care not to bend the sample
probe.
When the cleaning is performed, the detector counter
(number of cycles since last cleaning of waste
chamber) is automatically reset.