2-8 Sysmex XP-300
Safety Information
Revised February 2013
Decontamination
Warning!
Before decontaminating the instrument, be sure to
turn off the power supply and unplug the power cord.
This is necessary to avoid the risk of electric shock.
When cleaning the instrument, always wear protective
gloves and gown. Also, wash hands after
decontamination carefully with antiseptic solution first
and with soap afterwards. Do not open the instrument
for decontamination inside. This is executed only by
Service Technician.
Information
To ensure decontamination of the instrument outer
surfaces, clean the instrument surface at the end of
the daily work. This has to be executed in the
following three situations;
- Regularly, at the end of a daily work,
- Immediately, during contamination with potentially
infectious material, and
- In advance of repair or maintenance by the field
technical service representative
Wipe off the instrument surfaces using a cloth
soaked with a suitable decontamination solution.
Please use one-way cloths, e.g. made of paper or
cellulose. The cloth may be moistened in a way only
that no wetness may reach the inside of the
instrument.
The indicated residence time of the decontamination
solution shall be observed.
If required, you may afterwards remove normal
contaminations with commercial neutral detergent, in
case these could not be removed by the
decontaminant.
As a last step the instrument shall be dried with a dry
one-way cloth.