7-4 Sysmex XP-300
Sample Analysis
Revised February 2013
Number of the samples that can be analyzed with one
pack of reagent
CELLPACK: Approx. 600 samples/20 L
(large cubitainer)
Approx. 300 samples/10 L
(small cubitainer)
STROMATOLYSER-WH: Approx. 470 samples/500 mL (bottle)
(The above values are the result of continuous analyses performed in
one day in the whole blood mode. Depending on the use conditions
of the instrument, the result may differ.)
Replenishing reagent
• Make ready a new reagent and make sure that it has not passed
its expiration date. (For detail, refer to “12.16 Replace
reagent”.)
• We recommend preparing “Reagent Replenishment Record” in
which to enter date of replenishment, name of reagent, reagent
lot No., expiration date, name of person who replenished. Such
a record would come in handy.
Inspection of waste
If waste is collected in the trap chamber on the left side of the unit
and the waste tank (when provided), discard the waste.
• Use a reagent that has been left at room
temperature (15 - 30°C) for more than 24 hours.
• If CELLPACK that has just arrived is used, “Blank
Error” may occur.
• After replenishing a reagent, make sure that its
background count is low before starting sample
analysis.
• As to a reagent that may have frozen, handle it in
accordance with the precautions stated on the
Package Insert. Otherwise, there is a possibility
that proper analysis cannot be performed.
• When replacing the reagent container, take care
not to have dust adhere to the cubitainer spout kit.
• After unpacking, take care to prevent entry of dust,
dirt, bacteria, etc. which could impair proper
analysis.
Shelf life after first opening
CELLPACK 20 L 60 days
STROMATOLYSER-WH 500 mL 90 days