7-16 Sysmex XP-300
Sample Analysis
Revised July 2012
The analysis starts, the status display indicates [Aspirating].
When sample aspiration is completed, status display
[Aspirating] changes to [Running]. When [Running] is
displayed, the sample can be removed safely.
The analysis results for all parameters are displayed about 60
seconds after starting the analysis.
• Do not remove the sample from the sample probe
while [Aspirating] is displayed. If the sample is
removed from the sample probe while [Aspirating] is
displayed, a correct analysis result may not be
obtained.
• Several seconds after the buzzer sounds “beep,
beep” and [Running] appears on the screen, the
rinse cup lowers. Remove the sample tube by that
time.
• To remove the sample tube, lower it straight down.
Take care not to bend the sample probe.
The sample probe is automatically rinsed, so there is no
need to wipe it clean.