7-18 Sysmex XP-300
Sample Analysis
January 2012
Preparing analysis samples in PD mode (1:26 dilution)
(1) Clean a container such as Erlenmeyer flask, beaker, etc. with
CELLPACK and remove any dirt.
(2) Using a syringe, etc., take CELLPACK into a cleaned
container.
(3) Using a probe (500 μL), collect 500 μL of CELLPACK and
dispense it into the micro tube.
(4) Using a capillary tube, etc., collect 20 μL of blood and dispense
it into the micro tube.
(5) Attach the cap and mix well.
When preparing a 1:26 dilution sample, use the tools listed below:
Diluent (CELLPACK)
Micro tube (MT-40, etc.)
Capillary tube
Probe: 500 μL
A container, such as erlenmeyer flask or beaker
A syringe
Risk of Infection
When preparing PD-mode analysis samples, always
wear protective garments and gloves. After
completion of work, wash hands with disinfectant.
If your hands are contaminated by blood, etc.,
infection of bacteria can occur.
Caution!
A 1:26 dilution sample is prone to platelet
agglutination. So, analyze the sample within 30
minutes after blood dispensing and diluting. If diluent
is dispensed in advance, evaporation and dirt mixing
will result in errors in analysis values. Therefore
prepare one sample at a time.