Oscilloscope Reference Math arbitrary lters
Frequency response of the available smoothing lters
The following table lists the available smoothing lters:
File name Len gth Normalized bandwidth
Stop band attenuation dB
smooth3.t
3 0.1558 -9.4
smooth5.t
5
0.0903 -12
smooth10.t
10 0.0446 -12.9
smooth20.t
20 0.0224 -13.2
smooth50.t
50 0.00887 -13.2
smooth100.t
100 0.0045 -13.2
smooth200.t
200 0.0022 -13.2
xxx
Hi
lbert transform lter
The ideal Hilbert transform lter has a gain of one at all frequencies and shifts the phase of all frequencies
b
y90degrees. Thistypeoflter is one of the types that may be specied in the Remez Exchange
algorithm. This lter departs from its desired behavior in the frequency range of 0 to 0 .025 times
the sample rate and also in the range of about 0.475 to 0.5 times the sample rate. This type of lter
can be used to create quadrature signals over a wide frequency range. The lename for this lter is
HilbertTransform90PhaseShift.t.
862 DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals