Oscilloscope Reference MyScope menu
MyScope menu
Use the MyScope Menu to create a new MyScope control window, open or edit an existing one, display
the active MyScope window, or recall recently used ones.
Natural antilog
1. Inserts the text EXP( into the math expression.
2. Enter an argument to the function.
The n atural antilog function raises
e to the power of the argument.
Negative duty cycle
This timing measurement is the ratio of the negative pulse width to the signal period expressed as a
percentage. The duty cycle is measure d on the first cycle in the waveform or gated region.
Negative Duty Cycle = ((Negative W idth) ÷ Period) × 100%
Negat
ive overshoot
This
voltage measurement i s measured over the entire waveform or gated region and is expressed as:
Negative Overshoot = (Low – Minimum) × Amplitude × 100%
Negative width measurement
This timing measurement is the distance (time) between the mid reference (default 50%) amplitude
points of a negative pulse.
The measurement is made on the first pulse in the waveform or gated region.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 873