Gen II User Manual
12-13 – Honeywell Performance Chapter 12 Revision 2.20, 20
12.1.2.16. When the analyzer switches to the Acquisition mode, the header at the top of the
screen will indicate “Acquisition” and the data quality indications, ACCEPTABLE,
QUESTIONABLE and UNRELIABLE will be shown to the right of the P/Set
indication. The normal time required for acquiring each point is one minute. If data
quality dictates, the scan period may be extended until the required number of
acceptable data samples is acquired.
12.1.2.17. When acceptable data is acquired, the screen will automatically display the review
screen shown below. As indicated at the bottom of the screen, use the [] and []
keys to select data points. Press the [F0] “Retake” key to retake this power set point
or press [OK] to accept this data and continue.