4.3 Washing and Cleaning
HA-8190V 4-27
2 Prepare the washing solution for tubes.
a Add 5 mL of washing solution for tubes to a sample tube.
b Set the sample tube in port 1 of the washing rack.
c Load the washing rack onto the sampler.
See step 4 in “2.5.1. Preparing Samples” on page 2-21.
3 Wash the tubes.
a Touch or .
Tube washing will start.
The screen shown on the right will appear. Do not turn off the
power to the instrument while this message is displayed.
Washing solution
for tubes (5 mL)
12345678910
NOTE:
Be sure to use the washing rack. Use of other racks may
damage the instrument or seriously degrade the column,
requiring its replacement.
Start
REFERENCE:
To have the instrument sleep after tube washing:
Touch . When <Use> has been set for <Startup timer>, select the next startup day (see “3.3.2. Setting
the Startup Timer and Entering the Sleep Mode” on page 3-6).
Timer