4.3 Washing and Cleaning
HA-8190V 4-49
5 Complete column installation.
a Touch .
b When “Has the column been replaced?” appears,
touch .
• The previous column counter value will be maintained.
• Mechanical initialization will start.
• When initialization is complete, the message “Fluid runs at the
same rate as during measurement.” will appear.
• When fluid pumping is complete, the [Reagent replace menu]
screen will appear again.
c Touch to return to the standby screen.
Performing HbA1c calibration
Finish
REFERENCE:
The message “W-062 The front or maintenance cover is
open. appears if is touched with the front cover
opened. Be sure to close the front cover before touching
.
If is touched by mistake, the column counter will
be reset to “0000” and you will no longer know when to
replace the column.)
Cancel
OK
NOTE:
Be sure to perform HbA1c calibration after washing the optical unit cell.
See “2.8. HbA1c Calibration” on page 2-38.