Chapter 5 Troubleshooting
HA-8190V 5-6
W-044 Pressure low
Cause Air has entered the fluid pump.
Fluid is leaking from the eluent tubes or fluid pump.
Remedy a Touch to clear the warning. After measurement stops, perform steps b and c.
b Remove air from the tube by priming (see page 4-52: On the [Priming menu] screen, select
<1. Automatic>).
c If the warning persists, pump fluid (see page 4-52: On the [Priming menu] screen, select <2. Pump>).
Open the temperature control box cover while pumping fluid, and tighten the tube that is leaking.
W-045
W-046
W-047
Temperature is outside range.
Cause The room temperature is outside the measurement environment temperature range of between 10°C
and 30°C.
The temperature controlling unit did not operate properly.
Remedy a Touch to clear the warning.
b Adjust the room to a temperature between 10°C and 30°C.
c If the warning persists, turn off the power and contact your distributor.
W-050 Unloading side is full of racks. Remove racks.
Cause The rack unloading side of the sampler is full of sample racks with already measured samples.
Something is obstructing the rack detection sensor.
* This warning occurs when the optional side sampler is attached to the instrument.
Remedy a Touch to clear the warning.
b Remove the sample racks from the rack unloading side.
c Remove the obstruction from the front of the rack detection sensor (rear left of the unloading side of the
sampler).
W-051 HbA1c STD. solution has expired. Retry calibration with unexpired solution.
Cause The standard solution is beyond its expiration date.
Remedy a Touch to clear the warning.
b Retry HbA1c calibration using a new standard solution (see page 2-38).
W-052 No hemolysis washing solution. Replace bottle.
Cause The hemolysis washing solution has run out.
The fluid level detection sensor cord of the hemolysis washing solution bottle is disconnected from the
instrument.
Remedy a Touch to clear the warning.
b If the hemolysis washing solution has run out, replace the bottle with a new one (see page 4-7).
c Properly connect the fluid level detection sensor cord of the hemolysis washing solution bottle to the
WASH terminal on the rear panel (see page 1-24).
W-053
W-054
W-055
No Eluent A Replace pack.
No Eluent B Replace pack.
No Eluent CV Replace pack.
Cause Eluent A, B or CV has run out.
Remedy a Touch to clear the warning.
b Replace the indicated eluent pack with a new one (see page 4-3).
OK
OK