2.6 STAT Measurement
HA-8190V 2-29
b Set the sample container straight in the STAT port.
c Push the STAT port in the direction of the arrow near “PUSH” by
hand until it locks.
d Gently pull the STAT port forward to ensure the port is locked.
4 Reserve STAT measurement.
a Touch .
While the (STAT reservation icon) is flashing:
STAT measurement has been reserved.
The STAT measurement is held until normal measurement can
be interrupted.
When disappears:
The beeper will sound and the piercing nozzle will come
forward.
The sample in the STAT port will be aspirated and STAT
measurement will start.
The STAT measurement number will appear at the right side
of <STAT No.>.
“Results”
The obtained STAT measurement result will be displayed on
the screen and printed out.
See “2.9. Displayed and Printed Reports” on page 2-44.
NOTE:
Set the sample tube fully to the bottom of the STAT port so
that it stands straight. If the tube is tilted, it may cause
damage to the piercing nozzle.
Start
Keep your hands away from the STAT port while
STAT measurement is reserved. You may be
injured by the piercing nozzle, the sample tube
spinning unit or sample racks while they are
moving.
REFERENCE:
STAT measurement during HbA1c control measurement:
In control measurements, one sample container of control is
measured repeatedly according to the number of times you
have set on the screen. If STAT measurement is reserved
during control measurements, the STAT measurement starts
after the completion of the set number of measurements of the
sample in the same container.