Chapter 2 Measurement
HA-8190V 2-30
5 Remove the sample from the STAT port.
a Check that the STAT reservation icon has disappeared.
b Push the STAT port in the direction of the arrow near “PUSH” by hand until it unlocks, then slide it forward.
c Remove the sample from the STAT port.
Setting the next sample:
In step 5-c, the measured sample can be replaced with another sample to perform the next
STAT measurement. Be sure to slide the STAT port forward when replacing the sample. You
may be injured by the piercing nozzle or other parts, if you try to replace the sample when the
STAT port is still in the aspiration position.