Showing Data on the Additional Info Row
MC4 User Guide 91
Maximum - Minimum
Calculates the difference between the maximum and minimum
measurement value found after a measurement was started or it was
reset.
Cumulative Average
Calculates the cumulative average of measurements after a
measurement was started or it was reset.
Cumulative average means that, after the calculation is started,
every new measurement is included in the average calculation. Thus
the amount of measurements included in the cumulative average
increases all the time and the average value's change rate typically
gets smaller with time.
Internal Temperature
°
,
°F or
K
When an internal or an external pressure module is used in a
window, the internal temperature of the pressure module can be
selected to be shown.
Normal Reading
A
Normal Reading refers to the window's
. It is
available when the following Special Measurements reserve
window's main value: All Error Display Modes as well as
Percentage, Scaling and Deviation measurement.
Deviation Reference
Available when deviation measurement is active. This is the entered
reference value MC4 uses to calculate the measurement's deviation
from.
Redundant Measurement (Secondary Measurement)
B
The secondary measurement used during redundant measurement.
Redundant Reference
The alert limit that is entered when redundant measurement was
started. Available only when redundant measurement is active.
Difference A
Displays the main measurement during a difference measurement.