EasyManua.ls Logo

BEAMEX MC4 - Temperature Indicators and Recorders

BEAMEX MC4
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Examples of Instrument Calibration
136 MC4 User Guide
Temperature Indicators and Recorders
This procedure suits temperature indicators/recorders no matter if
the sensor is an RTD or a thermocouple. The instrument's input sig-
nal is simulated using MC4. The instrument's output (reading) is
manually entered into MC4.
Preparations
1. Connect suitable terminals in MC4 to the input connectors of
the indicator/recorder.
2. If the sensor to be simulated is a thermocouple, select one of
the following reference junction methods:
MC4's internal reference junction is used.
The reference junction temperature is measured exter-
nally using an RTD connected to the RTD terminals.
The reference junction temperature is fixed to 0°C.
The reference junction temperature is manually en-
tered.
Additional information about Reference Junction com-
pensation methods can be found in chapters
Internal Reference Junction on page 113 and
External Reference Junction on page 114.
3. If the sensor to be simu-
lated is an RTD, the used
wiring system depends on
the instrument to be cali-
brated. Remember to only
use the two leftmost RTD
terminals in MC4.
4. Select the instrument in
MC4 as presented in
chapter Selecting the In-
strument to Be Calibrat-
ed on page 129 or create
a new instrument as de-
scribed in chapter In-
strument List Window
Menu on page 152.
6. Test the connections be-
fore starting calibration,
i.e. see if the input and
output signal are "live" in
MC4.

Table of Contents

Related product manuals