Utilities Menu
100 MC4 User Guide
Notes.
All changes in general settings become valid immediately except
when changing the language. To activate the new language, shut off
and restart MC4.
Disable auto-off delay during long-time tasks (e.g. leak/stability test).
Otherwise the auto-off functionality interrupts the task.
When calibrator's Auto-off delay is in use (value other than zero), a
warning dialog pops up 30 seconds before the auto power off
executes. In the dialog the Function Keys have the following options:
F1/
Cancel
Cancels the auto power off feature for the moment, but it stays
active even during this session.
F2/
Stop
Disables the auto power off feature for this session. It will
reactivate when the calibrator is switched on the next time.
F3/
Ok
Immediately shuts down the calibrator.
Calibrator Adjustment
Opens a window to start the recalibration and adjustment of MC4.
See Appendix 2 for instructions on how to recalibrate MC4.
E n t e r p a s s w o r d t o e n a b l e
c a l i b r a t o r a d j u s t m e n t s
C A L I B R A T O R A D J U S T M E N T
1 6 . 0 8 . 2 0 0 7 0 9 : 4 5 : 2 2
I n t . P r e s s u r e ( I P M 2 0 C )
*
3 6 0
C a l i b r a t i o n P e r i o d
B e a m e x C a l i b r a t e d b y
E l . M e a s u r e ( E L - M E A S U R E )
S t o p C l o s e