Additional Calibration Related Information
182 MC4 User Guide
Combining Calibration Repeats
There is a check box entitled
Combine with Previous Calibration
Results
on the
Calibration Info
page. If checked, the latest cali-
bration will be combined with the previous result as one calibration
event when it is saved. All combined results are received in calibra-
tion software (such as CMX) as one calibration event containing
several repeats. The check box is automatically set if there is less
than three days between the current and previous calibration.
If you have defined
Number Of Calibration Repeats
for the in-
strument, then MC4 opens a window after you have saved calibra-
tion results. The window informs you of how many repeats are done
and how many repeats are supposed to be done. More of this fea-
ture in chapter General Data Page on page 172.
Results and Memory Usage
If you are calibrating instruments using off-line method then please
keep in mind this: When you upload the results from MC4 to the cal-
ibration software, the results are by default also left in MC4's
memory. Delete results to free memory in MC4 as described in
chapter Instrument List Window Menu on page 152.