66 MC4 User Guide
Manually Keyed Values
MC4 supports the use of manually entered values. They are mainly
of use when calibrating instruments (see part D), but they may also
be used in Basic Mode. The following example presents one use of
manually entered values:
This suits instruments that have an input signal not supported by
MC4, but the output can be measured by MC4. Actually, the input
and output may be interchanged too as long as the one of the
signals is measured and the other is entered manually.
Window settings:
Function:
Manually Keyed Value
Window settings, e.g.:
Function:
Current Meas., Supply On
or
Current Meas., Supply Off
Open the Tools menu for window . Select
Display Mode /
Specials
. Choose one of the available error calculation modes.
More of error calculations in chapter Display Mode and Special
Measurements on page 81. Enter the input and output ranges in the
opened window.
Connect the instrument's current output (or whatever measurement
you chose for window ) to MC4 and manually enter the value of
the other signal to window . Then you will be able to view the error
of the instrument's input/output relation.
Note.
MC4 deduces the resolution for the manually keyed value based on
entered numbers. To ensure valid resolution, enter e.g. trailing zeros
accordingly.