Calibrator Adjustment
MC4 User Guide 101
Custom Test Point Sets
MC4 has a comprehensive set of pre-entered test point sets (steps).
If however they are not suited for your needs, you can create your
own test points as presented here.
In Basic Mode, custom test point definition can be started from the
following utilities found in the
Tools
menu for generation/simulation
Functions:
• Stepping presented on page 73 and
• Manual Stepping presented on page 78.
One of the utilities' configuration fields is
Step Definition
combo.
The last option in the combo is
Create New
. It opens the custom
test point tool. If a custom set already exists, selecting it in the
Step
Definition
combo also opens the custom test point tool.
In Calibration Mode, custom test point definition can be started
while editing instrument data. See:
• Procedure Data Page presented on page 165.
In this case, the custom test point tool opens from
Calibration
Points
combo.
Give a name to the Custom test
point set. The
Scaled Range
initially has the same range as
defined for the Stepping or Manual
Stepping from where the custom
test step definition was invoked.
You may edit the
Scaled Range
according to you own needs during
step definition, e.g. scale the steps
to percentage values.
Enter the point values. Press down
arrow ( ) to add a point below the
last point.