EasyManua.ls Logo

BEAMEX MC4 - Electrical Generation, Sensor Measurement and Simulation; MV Measurement (T;C-Terminals); MV Generation (T;C-Terminals)

BEAMEX MC4
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix 1, Technical Data
MC4 User Guide 187
Electrical Generation, Sensor Measurement and Simulation
Available in Temperature/Generate section.
mV Measurement (T/C-Terminals)
Range Resolution 1 Year Uncertainty
(1
-25 ... 150 mV 0.001mV 0.02 % RDG + 4 µV
Feature
Specification
Temperature
coefficient
Add 0.0015 % RDG /°C to spec. outside 18  28°C
Add 0.0008 % RDG /°F to spec. outside 64.4  82.4°F
Input impedance >10 Mohm
Supported units V, mV, µV
Display update rate 3 / second
mV Generation (T/C-Terminals)
Range Resolution 1 Year Uncertainty
(1
-25 ... 150 mV 0.001 mV 0.02 % RDG + 4 µV
Feature
Specification
Temperature
coefficient
Add 0.0015 % RDG /°C to spec. outside 18  28°C
Add 0.0008 % RDG /°F to spec. outside 64.4  82.4°F
Max. load current 5 mA
Load effect < 5µV/mA
Supported units V, mV, µV
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period. (k=2)

Table of Contents