Appendix 1, Technical Data
MC4 User Guide 187
Electrical Generation, Sensor Measurement and Simulation
Available in Temperature/Generate section.
mV Measurement (T/C-Terminals)
Range Resolution 1 Year Uncertainty
(1
-25 ... 150 mV 0.001mV 0.02 % RDG + 4 µV
Feature
Specification
Temperature
coefficient
Add 0.0015 % RDG /°C to spec. outside 18 28°C
Add 0.0008 % RDG /°F to spec. outside 64.4 82.4°F
Input impedance >10 Mohm
Supported units V, mV, µV
Display update rate 3 / second
mV Generation (T/C-Terminals)
Range Resolution 1 Year Uncertainty
(1
-25 ... 150 mV 0.001 mV 0.02 % RDG + 4 µV
Feature
Specification
Temperature
coefficient
Add 0.0015 % RDG /°C to spec. outside 18 28°C
Add 0.0008 % RDG /°F to spec. outside 64.4 82.4°F
Max. load current 5 mA
Load effect < 5µV/mA
Supported units V, mV, µV
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period. (k=2)