About Instrument Calibration
MC4 User Guide 139
Calibration
1. In the calibration window: press F3/
Start
.
2. Set the temperature of the bath/dry block to the required cali-
bration point. Wait until the temperature stabilizes.
3. If the temperature of the reference sensor is entered manual-
ly, enter the value.
4. How a calibration point is accepted depends on the procedure
settings of the instrument at hand:
• If
Automatic Calibration Point Acceptance
is set
(checked) and applicable, the input and output values of
the current point are saved automatically. Automatic
acceptance settings are presented in chapters Proce-
dure Data Page on page 165 and Accepting Calibra-
tion Points Automatically on page 178.
• If
Automatic Calibration Point Acceptance
is not
set/applicable, accept the point using F3/
Accept
Func-
tion Key. Note that this functionality is also available
when
Automatic Calibration Point Acceptance
is
set. This allows you to save results when the input
and/or output signals are permanently unstable.
5. Proceed to next calibration point (step 2) if applicable. When
the calibration run is ready, save or reject the results as de-
scribed in chapter A Calibration Procedure Using MC4 on
page 130.
6. Either do another calibration run, adjust the instrument or end
the calibration by selecting F1/
Menu
,
Close
.