EasyManua.ls Logo

BEAMEX MC4 - Page 204

BEAMEX MC4
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix 1, Technical Data
192 MC4 User Guide
Ni 120
Range Resolution 1 Year Uncertainty
(1
-80  260 °C 0.01 °C 4-wire measurement:
0.06 °C
Simulation
(2
:
0.12 °C
Cu 10
Range Resolution 1 Year Uncertainty
(1
-200  260 °C 0.01 °C 4-wire measurement:
0.2 °C
Simulation
(2
:
0.8 °C
RTD Types Available as Standard
Pt50 (385) Pt1000 (385) Ni100 (618)
Pt100 (385) Pt100 (3926) Ni120 (672)
Pt200 (385) Pt100 (391) Cu10 (427)
Pt400 (385) Pt100 (375)
Pt500 (385) Pt100 (3923)
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period. (k=2)
2)
Specification valid with an excitation current
>0.2 mA (0...400 ohm),
>0.1 mA (400...4000 ohm)

Table of Contents