Appendix 1, Technical Data
192 MC4 User Guide
Ni 120
Range Resolution 1 Year Uncertainty
(1
-80 260 °C 0.01 °C 4-wire measurement:
0.06 °C
Simulation
(2
:
0.12 °C
Cu 10
Range Resolution 1 Year Uncertainty
(1
-200 260 °C 0.01 °C 4-wire measurement:
0.2 °C
Simulation
(2
:
0.8 °C
RTD Types Available as Standard
• Pt50 (385) • Pt1000 (385) • Ni100 (618)
• Pt100 (385) • Pt100 (3926) • Ni120 (672)
• Pt200 (385) • Pt100 (391) • Cu10 (427)
• Pt400 (385) • Pt100 (375)
• Pt500 (385) • Pt100 (3923)
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period. (k=2)
2)
Specification valid with an excitation current
>0.2 mA (0...400 ohm),
>0.1 mA (400...4000 ohm)