Generating/Simulating
62 MC4 User Guide
Notes.
In RTD simulation MC4 monitors the resistance measurement
current. If the current is too high, it cannot simulate the right
resistance value. In that case it shows the message "
High Cur
".
Accurate operation of the simulation electronics requires that the
current supplied by the instrument under test does not vary rapidly.
The simulation result is not accurate if the instrument under test
uses AC current. If the instrument under test uses pulsed
measurement current, it should wait a few milliseconds before
starting the measurement after setting the current.
Next
Temperature Measurement (RTD) on page 50.
Thermocouple Simulation on page 63.
Resistance Simulation on page 60.
Tools Menu on page 67.
Stepping on page 73
Ramping on page 76.
Calibration, see Part D.