6.2.5 SA
Summary: This command moves Axis1 to the desired Target Position. This is an
old style SR80 Command. It is not recommended for new applications.
Note: You must not attempt to move to a new target position while the Axis is in
motion. Query the motion status by using the GetAxisStatus Command. This
command is replaced by the SetTarget Command and should be not used for new
applications.
SR-80 command format compliance.
Reference Command: SetTarget, GetAxisStatus. SetActiveAxis
S A
Space
TargetNumber
EOM
Input Operand: TargetNumber. This is a value between 1 and the number of
Targets/Filters/Sources on Axis1. Any value out of range will result in an invalid
operand.
Return Values: None
605-7040 Operation Manual