5.2 Maintenance Concept
I-Level maintenance technicians are authorized to perform the following maintenance
tasks on the SR-800R:
Troubleshooting (Section 5.3)
Preventive/Scheduled Maintenance Tasks (Section 5.4)
5.3 Troubleshooting
This section includes description of the troubleshooting principle followed by a list of
all-possible failure symptoms and instructions for locating them.
Table 5-1 details how to troubleshoot failures that may occur while using the
SR-800R.
Table 5-1: Troubleshooting
Message
No.
Failure Message
Probable Cause of Failure/
Action
25 "ERROR: Could not open the Remote
Communications Channel."
Inspect the communication
cables.
Restart the controller.
30 "ERROR: IO Port Short-Circuit! Please check the
short-circuit in the IO output accessories or cables
and restart the system."
Fix the shorted circuit.
Restart the controller.
31 "ERROR: Power Amplifier Short Circuit. The
Power Amplifier has been shut down. Possible
causes are: -Electrical short circuit inside the
Black Body head. -Short circuit in the
interconnect cable. Please fix the problem and
restart the system."
Fix the shorted circuit.
Restart the controller.
32 "WARNING: The Control Master cannot reach
the required differential temperature (dT),
because the absolute Te temperature value cannot
be above XX
1
degrees. You must select a lower
dT value."
Re-enter new dT values.
1
: A parametric value; taken from the actual screens.