3.2.3.5.4 MRTD Test - Run Window
1. From the MRTD Test Window, press . The window characteristics change
to the following:
Figure 3-25: MRTD Test Run
- The Run button is temporarily replaced with the Wait button and
then to the Next button.
- The Abort button becomes active.
- The Setup button becomes non-active.
- The message Step 1 is displayed on the top of the window.
- The Setpoint field shows the DT Start temp of Step 1 and the current field shows
the actual temperature (you may clear the temperature values by clearing the Current
and Setpoint selections).
- The MRTD Test is initiated while the Status field displays the
Not
Ready indication. The Current target color becomes red to indicate a transient
situation for the temperature and target of Step 1 on the MRTD test.
- When the target has reached its position and the BB temperature has stabilized, the
Status field displays the
Ready indication, and the red target color turns
yellow.
2. Observe the target on the FILR monitor and change the temperature by using the
Temperature up/down arrows. Wait until the temperature has stabilized, and the
Ready indication is re-displayed in the Status field.
3. When you have reached the "Just Resolved" point, proceed to the next step, and
press .
Now the controller changes the Setpoint temp to the negative value of the Step 1
DT Start. Wait for stabilization.
4. Upon completion of the positive and negative dT, pressing
will change
the Target and the dT to the next step.