3.2.3.6.5 Touch Panel Calibration
The Touch Panel Calibration option allows you to calibrate the SR-800R Controller
Soft-Touch Panel.
NOTE
Use a stylus to calibrate the Touch Panel.
1. From the Options window, press
.
2. Follow the prompts. Hold the stylus for at least two seconds on each cross-sign.
Upon completion, the Options window re-opens.
3.2.3.6.6 Temperature
The Temperature option allows you to set: the temperature display resolution, the
arrow increments and the temperature stabilization attributes.
1. From the Options window, press
. Pending on the blackbody
working mode, the following window opens:
Figure 3-46: Absolute Temperature Setting
605-7040 Operation Manual
Rev D 3-44
June 2013