8.1 Deliverables
SR-800R Blackbody, consisting of:
Temperature Controller
Blackbody
Connection cables: Power cable, Head cable, RS-232 cable
Hood (optional)
Refrigerated Circulator, consisting of:
Julabo Refrigerator unit
Cooling fluid
Two insulated pipes
Power cables
100/220 VAC transformer (supplied only with the 100 V option)
Julabo Circulation System Operating Manual
8.2 Safety Precautions
Read all safety instructions stated in the Julabo Circulation System Operating
Manual.
Hazards are posted throughout this manual and the Julabo Circulation System
Operating Manual. Follow these hazards to avoid possible injuries to personnel
and damage to equipment. Always follow your local safety precautions.
Note that the refrigerator’s total power consumption is 3,100W, use a suitable
plug and power outlet.
Exercise care never to spill cooling fluid on the Blackbody.
Before starting the circulator, check the cooling fluid level. It must be higher than
the heater/refrigerator coils.
After running the circulator for a short time, check the cooling fluid level again. It
may drop as the fluid fills the circulation piping. If it does, refill to the required
level.
If you intend to operate the SR-800R at low temperatures or sub-freezing
temperatures, make sure that the Blackbody is placed in a humidity-free chamber;
otherwise, ice/condensing will form on emitter plate.