3.2.3.6.2 Settings
The Settings window allows you to set some defaults parameters of the SR-800R
Controller.
From the Options window, press
. The following window opens:
Figure 3-39: Settings
1. To Set the PowerUp Configuration:
a. If you want the SR-800R to boot each time you turn the system ON with the
last settings before it was turned OFF, select the Recall Last option.
The affected settings are:
- The last Set Temperature of the Absolute/Differential.
- The last Temperature Scale (Celsius/Kelvin).
- The Radiometric Offset (On/Off).
- The last status of the IO_Out Ports (On/Off).
- The last selected target.
b. If you want the SR-800R to boot with the system defaults settings, select the
Default option.
The system defaults are:
- Absolute temperature of 25°C and Differential temperature of 0°C.
- The Temperature Scale is Celsius.
- The Radiometric Offset is Off.
- The IO_Out Ports are Off.
605-7040 Operation Manual
Rev D 3-36
June 2013