EasyManua.ls Logo

CI Systems ControlMaster SR-800R - Page 72

CI Systems ControlMaster SR-800R
185 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CI Systems SR-800R
2. To Set the Buzzer Configuration:
a. If you want to enable or disable the Touch Panel Buzzer you can
select/deselect the Touch Panel checkbox.
b. If you want to enable or disable an audible beep when the temperature is
stable you can select/deselect the Temp Stable checkbox.
3. To Set the Main Axis:
You can select the motion control axis that is controlled via the Main window (the
TARGET button) by selecting it from the available axis list of the Main Window
Axis.
4. To Set the Blackbody Head Type of the SR-800R: use the drop-down menu to
select the Head type of the blackbody. There are 2 types that can be defined:
Abs&Diff and Abs only. Abs Only is a must if the target sensor is disconnected
or not exist.
NOTE
Blackbodies that operate in Abs&Diff mode require that the
Head be connected to Target Sensor. Otherwise, the system will
not operate correctly.
A factory supplied Differential head can always operate in Abs
mode but an Abs Only factory supplied head cannot operate in
Differential mode.
605-7040 Operation Manual
Rev D 3-37
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents