CI Systems SR-800R
2. Remove the two Phillips screws securing the Sensor Locker to the Blackbody.
3. Remove the Sensor Locker (see Figure 7-8).
NOTE
The inserted sensor is held in place by way of a built-in magnet;
therefore, a reasonable force is required to separate the sensor
from its port.
4. Remove the Emitter Sensor using the Sensor Extractor Jig (the jig is supplied
with the new sensors):
Figure 7-8: Emitter Sensor Replacement (2" Blackbodies)
Sensor Extractor Jig
Sensor Locker
2
4
Fastening
Screw (x2)
3
5. Store the sensor in its package to avoid damage.
605-7040 Operation Manual
Rev D 7-8
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com