EasyManua.ls Logo

CI Systems ControlMaster SR-800R - Target Sensor Replacement

CI Systems ControlMaster SR-800R
185 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CI Systems SR-800R
7.3 Target Sensor Replacement
This section includes step-by-step procedures for replacing the Target sensor in the
following blackbody types:
4 ÷ 20" Blackbodies (see Section 7.3.1)
2" Blackbodies (see Section 7.3.2)
7.3.1 4 ÷ 20" Blackbodies
7.3.1.1 Removal
1. Disconnect the Sensor male connector from the Blackbody:
Figure 7-9: Sensor Cable-Connector (typical)
7"
12 ÷ 20"
4"
2. Hold the sensor itself and carefully pull it out (do not pull it by its cable).
NOTE
The inserted sensor is secured to the emitter by way of a built-
in magnet. Therefore, a reasonable force is required to separate
the sensor from the emitter.
605-7040 Operation Manual
Rev D 7-10
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com