CI Systems SR-800R
7.3 Target Sensor Replacement
This section includes step-by-step procedures for replacing the Target sensor in the
following blackbody types:
4 ÷ 20" Blackbodies (see Section 7.3.1)
2" Blackbodies (see Section 7.3.2)
7.3.1 4 ÷ 20" Blackbodies
7.3.1.1 Removal
1. Disconnect the Sensor male connector from the Blackbody:
Figure 7-9: Sensor Cable-Connector (typical)
7"
12 ÷ 20"
4"
2. Hold the sensor itself and carefully pull it out (do not pull it by its cable).
NOTE
The inserted sensor is secured to the emitter by way of a built-
in magnet. Therefore, a reasonable force is required to separate
the sensor from the emitter.
605-7040 Operation Manual
Rev D 7-10
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com