8.4.4 Cable-Connection
1. Connect the Head cable between the Blackbody and the SR-800R Controller (see
Figure 8-1).
2. Connect the Julabo Circulation System cables (see Figure 8-9):
a. Connect the RS-232 Cable between the SR-800R Controller and the Julabo
SERIAL cable-connector.
Note that the Circulator provides three possibilities for the setpoint selection,
i.e., Temperature Setting.
At CI-Systems, the Circulator is set to work in the RS-232 mode with the
following communication setup (for reference only):
• Baud - 9600
• Parity - None
• Stop Bit - 1
• Data Length - 8 Bit
• Handshake - None
NOTE
Depending on the SR-800R Controller temperature setpoint, the
only way to control the Julabo Circulation System temperature
is by the SR-800R Controller and via the RS-232
communication port. Therefore, the SR-800R Controller RS-
232 port cannot be used for other purposes.
b. Connect the Julabo Control Cable between the Julabo Controller and the
ALARM cable-connector.
c. Connect the Julabo Controller Power Cable to the Circulator Power cable-
connector.
d. Connect the Main Power Cable to the Mains (either directly or via a 100
VAC transformer).
605-7040 Operation Manual
Rev D 8-13